Coil current monitors

From LTX Wiki
Jump to: navigation, search

Currents in the equilibrium field coils are measured by Rogowski coils wrapped around the field coil leads at various points in the LTX test cell (L210/211), mostly near the Hoffman boxes, and also be integrated current sensors built into the power supplies.

Acquisition

Main Rogowski Coils

The coil current Rogowskis are commercial parts purchased from Rocoil, and are documented as having an instrument gain of 1.07x10-7 V/A.

Calibration testing in Jan. 2019 found that this nominal gain is generally accurate. Most Rogowskis that were tested fell within ~1% of the nominal gain. The largest deviation measured was the Internal coil Rogowski, which appeared to have a 5% higher gain when measured with a second wire run through the Rogowski next to the main coil lead. The 5% difference was deemed small enough to not bother conducting further experiments to test whether they persisted when driving current through the main coil lead itself.

Coil Rogowski Integrator Digitizer
Location Wrappings Gain (Vs/A) Chassis Channel Gain (1/s) Chassis Channel
TF On red lead at TF terminal plate 3 1.07e-7 1 30 505.5 Neos 90
OH On OH bus above NBI 1 1.07e-7 1 29 500.9 Neos 88
Blue at Hoffman box 3 1.07e-7 1 13 1448 Neos 81
Red at Hoffman box 2 1.07e-7 1 14 2952(?) Neos 82
Orange (top) at Hoffman box 3 1.07e-7 1 12 4844 Neos 84
Orange (bottom) at Hoffman box 4 1.07e-7 1 15 2287 Neos 85
Yellow Inside yellow cap bank rack 1 1.07e-7 1 32 503.4 Neos 83
Green above green cap bank rack 4 1.07e-7 1 31 2289 Neos 87
Internal at Hoffman box 3 1.07e-7 1 16 2887 Neos 86

Extra/Undocumented Rogowski Coils

Name Rogowski Integrator Digitizer
Location Wrappings Gain (Vs/A) Chassis Channel Gain (1/s) Chassis Channel
"OH rog" ? ? ? ? ? ? Neos 89
"Yellow current" ? ? ? ? ? ? Neos 80
"Green current" ? ? ? ? ? ? Neos 79

Power Supply Monitors

  • Blue: Power supply T1 is recorded in Helios input 44.
  • Internal: Power supply T2 is recorded in Helios input 45.
  • Orange (T + B): Power supply T3 is recorded in Helios input 46.
  • Green: Power supply T4 is recorded in Helios input 47.
  • Red: Power supply DF is recorded in Helios input 64.
  • TF: Power supply IF is recorded in Helios input 48.

Related Pages

Equilibrium field coils - lists parameters of the eight field coils Coil current monitor testing - brief calibration experiments to verify the coil current monitors' accuracy

Diagnostics

Magnetics (Local): Reentrant array | Toroidal array | Saddle coils | Shell eddy sensors
Magnetics (Areal): Poloidal flux loops | Loop voltage | Diamagnetic loop | IP Rogowski
Spectroscopy: Thomson scattering | ChERS | X-ray spectroscopy | Spectrometry | Vacuum UV spectroscopy
Microwaves: Reflectometry | Interferometry
Physical Probes: HFS edge probes | LFS SOL probes | RFEA
Operational Diagnostics: Ion gauges | Coil current monitors | Thermocouples | RTDs